Parameter
5
1
Example:
Motor temperature
Parameter
ASCII
0
HEX
30 30 31 34

5.4.10 ReadStatus

Function: Reads various settings (Remote mode, TMS function, Emergency vent valve).
Transmission frame:
PC->SIM Stx
SIM->PC
PC->SIM
SIM->PC Stx
Parameter
1
2
*1
3
4
*1
*1 Valid only with TMS specification. Do not set "ENABLE" without TMS specification.
MT-89E-001-C
Page 102
STP-iS2207 Series Turbomolecular Pump
Item
Motor temperature
(Unit: ºC)
: 0014
= 20 ºC (68 ºF)
hex
1
0
1
4
0
0
1
?
f
0
0
1
Sp
f
Item
Remote mode setting
TMS function setting
[System reservation]
Emergency vent valve setting
Data format
16-bits hexadecimal
coded ASCII
Etx LRC
Ack
Parameter 1 to 4
Etx LRC
Data format
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
Remark
Ack
Remarks
Refer to Table 27
00
: ENABLE
hex
Excluding 00
: DISABLE
hex
00
: ENABLE
hex
Excluding 00
: DISABLE
hex