Table 20 shows ASCII characters being used in the transmission control, error control and handshake
in the application layer.
5
Transmission layer
Application layer
5.3.2

Standard transmission frame (in the RS232/RS485 single point connection)

The transmission frame has a single block or multiple transmission blocks. The transmission block
consists of a start control character, data block No. (3 digits), a message (up to 255 characters), an
end control character, and a checksum (LRC). The following table shows the transmission frame
where the message transmission character string is C
Transmission frame when a message is below 255 characters (n<=255):
"Stx" and "Etx" are used as a start and an end character of the transmission frame, respectively.
MT-89E-001-C
Page 78
STP-iS2207 Series Turbomolecular Pump
ASCII
HEX
character
code
Stx
02
Etx
03
Etb
17
Ack
06
Nak
15
@
40
#
23
!
21
Table 20 - Transmission control characters
1
2
3
ASCII
Stx
0
0
Function
Transmission block start character
Transmission frame end character
Transmission block end character
Acknowledgment response
Non-acknowledgment response
Network frame ID character
Acknowledgment response
Non-acknowledgment response
.
n
4
5
5+n
1
C
C
1
n
5+n+1 5+n+2
Etx
LRC