7.6.2
Indication of "FAILURE" LED (green)
7
If an abnormality/error is detected by self test during the power ON operation, "FAILURE" LED flashes
in green. This is the indication of STP pump or control unit failure. Contact Service Office.
7.6.3

No Indication of the "FAILURE" LED

No.
Symptom
1
The "POWER"
LED does not
illuminate.
2
The
"ROTATION"
LED does not
flash after
performing the
start operation.
3
Insufficient
ultimate
pressure.
4
Abnormal noise
is generated
while the pump
is rotating
MT-89E-001-C
Page 144
STP-iS2207 Series Turbomolecular Pump
Probable cause
Incorrect connection of
power cable
Abnormal power voltage Set the input voltage to the rated voltage
Power failure
Insufficient cooling
Failure of the "Input
Operation Port" setting.
Failure of the
backing-pump.
Leakage of the piping
system.
Residual molecules.
External vibration or
impact
Table 43 - Troubleshooting with no indication of the "FAILURE" LED
Countermeasures
Connect the power cable correctly.
±10 %.
Check whether or not a power failure
has occurred.
Make sure that the STP pump is cooled. 2.3
Set the input operation port correctly.
Check the capacity and starting state
(use a backing-pump having more
capacity that we specified).
Check the piping system for leakage.
If the main composition is H
perform baking; if it is other gases, clean
the inside of the vacuum equipment (If
gases remain inside the STP pump,
contact Service office when it is needed
to be cleaned).
Remove external vibration so as not to
transmit it to the STP pump.
Referred
Section
3.4.5
3.4.6
2.4
7.1.1
4.9
3.5
3.5
or H
O,
4.3
2
2
3.1.1