5.4.17 Setoptionfunc - Edwards STP-iS2207 Series Instruction Manual

Turbomolecular pump
Table of Contents

Advertisement

5.4.17 SetOptionFunc

5
Function: Changes the setting value of the input operation port, warning function, and serial
communication time out.
There is an endurance limit of setting changes (no more than 24 times per day for about ten years).
If this limit is exceeded, it not only may become impossible to change the setting but can also cause
pump failure. Do not make setting changes in excess of this endurance limit.
Transmission frame:
PC->SIM Stx
SIM->PC
PC->SIM
SIM->PC Stx
*1 The HEX code of ASCII character ' = ' is "3D".
Parameter
1
2
*1
3
4
5
6
7
8
9
10
MT-89E-001-C
Page 114
STP-iS2207 Series Turbomolecular Pump
*1
0
0
1
Sp
=
Parameter 1 to 13
0
0
1
#
Etx LRC
Item
Input operation port setting
TMS Option
Enable/Disable setting
[System reservation]
Second Damage Limit Option
Enable/Disable setting
First Damage Limit Warning
Enable/Disable setting
Pump Runtime Over Warning
Enable/Disable setting
Pump Runtime Over Warning
Hours setting (×100 hours)
Imbalance Warning
Enable/Disable setting
Pump Overload Warning
Enable/Disable setting
Pump Overload Warning
Motor current setting
CAUTION
Etx LRC
Ack
Data format
8-bits hexadecimal coded
ASCII
8-bits hexadecimal coded
ASCII
48-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
32-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
16-bits hexadecimal
coded ASCII
Ack
Remarks
Default 01
hex
(I/O REMOTE)
Default FF
hex
(DISABLE)
*2
Default 00
hex
(ENABLE)
Default 00
hex
(ENABLE)
Default FF
hex
(DISABLE)
Default
00000000
hex
Default 00
hex
(ENABLE)
Default FF
hex
(DISABLE)
Default 03E8
hex
(100.0%)

Hide quick links: