Assembly Replacement Procedures

Top Brace

Top Brace

Removal

1. Remove the instrument outer case. Refer to the
2. Refer to
Figure 7-4
Top Brace Removal

Replacement

1. Refer to
168
procedure.
Figure
7-4. To remove the top brace (1), use the T-10 driver to
remove the twelve screws (2) (0515-0372) attaching the top brace to the
chassis and the fourteen screws (3) (0515-1227) attaching the top brace to the
boards.
Figure
7-4. To replace the top brace, place it into the correct position
and attach the appropriate screws. Torque to 9 inch-pounds.