Replacement Procedure

Outer Cover Removal

Outer Cover Removal

Tools Required

Procedure

Refer to Figure 6-1 for this procedure.
Step 1. Disconnect the power cable from the E5061A/E5062A.
Step 2. Remove the four TORX T15 screws (item 6) fastening the four rear foot
Step 3. Remove the four TORX T15 screws (item 9) fastening the cover.
Step 4. Remove the two TORX T20 screws (item 1,4) fastening the side strap handles.
Step 5. Remove the four bottom feet (item 3, 5).
Step 6. Remove the three TORX T15 screws (item 8) from the bottom.
Step 7. Slide the outer cover and remove it carefully.
Figure 6-1
Outer Cover Removal