Configure Settings - Primes MicroSpotMonitor Plus HighBrilliance Original Instructions Manual

Table of Contents

Advertisement

MSM+ HB
These two values are displayed in the two coordinate windows of the Beam Pointing tool
9.9.1

Configure settings

1.
Click on the Settings tab.
2.
Maintain the options according to the
explanations in the following table.
Option
z-Position in mm
Number of measurements
z-Axis position (z1 / z2)
Tab. 9.11:
Options in the Device control > Settings tab of the Beam pointing measuring mode
Revision 00 EN - 02/2022
Explanation
Use one of the following options to specify the desired initial z-position
of the camera housing for the pre-caustic measurement:
X Enter a value in the input field.
X Use the slider below the input field.
It is recommended to start the pre-caustic 1 or 2 Rayleighlength above
focus.
X Enter the number of measurements. The default setting is 15 mea-
surements.
With this input you determine the number of measuring points that are
displayed in the coordinate windows of the Beam Pointing Stability
tool.
For an unlimitted and recorded measurement enter the value „0".
The measuring positions z1 and z2 determined during the pre-caustic
measurement are displayed here.
85

Advertisement

Table of Contents
loading

This manual is also suitable for:

Msm plus hb10Msm plus hb20