Configure Advanced Settings - Primes MicroSpotMonitor Plus HighBrilliance Original Instructions Manual

Table of Contents

Advertisement

MSM+ HB
9.7.2

Configure advanced settings

1.
Click on the Advanced tab.
2.
Maintain the options according to the
explanations in in the following table..
Option
Save device settings
# Pixel in x/y *
Averaging *
Number of averaged planes *
Low power mode
Calibrated wavelength(s) in nm *
Tab. 9.8:
Options in the Device control > Advanced tab of the Manual caustic measuring mode
Revision 00 EN - 02/2022
Explanation
All options marked with an asterisk in the Device control menu can be
saved in the EEPROM of the device.
These and other options for saving / loading configurations are descri-
bed in chapter 9.4.4 on page 53.
This option enables setting the resolution of the CCD sensor:
X Enter the number of pixels in the x-direction / y-direction in the input
field on the left / right.
If this option is enabled, different algorithms for averaging a plane
measurement can be selected in the drop-down list below. Averaging
over several measurements can be useful, for example, when measuring
a laser with significant power fluctuations.
1.
Set the check mark to enable the option
2.
Select an algorithm from the drop-down list:
Arithmetic mean: The measured values for each pixel are added
together and divided by the number of planes.
Max. intensity per pixel: The values from all the measurements are
compared for each pixel and only the maximum value for each one
is displayed.
Max. lines: The values from all the measurements are compared for
each line (meaning the line issuing from the measuring device in the
x-direction for example) and only the maximum value for each one is
displayed.
X With the Averaging option enabled, enter the number of plane
measurements for averaging.
We recommend enabling this option for relatively low power laser be-
ams.
If this option is enabled, the following initial values are selected during
beam search
Filter level of the filter wheel: 0 (instead of 1 otherwise)
Integration time: 10 ms (instead of 1 ms otherwise).
X Set the check mark to enable the option.
X Select the calibrated wavelength in the drop-down list.
The values displayed correspond to the calibration points of the
measuring objective. The correct value should be selected before each
measurement to enable high measurement accuracy.
75

Advertisement

Table of Contents
loading

This manual is also suitable for:

Msm plus hb10Msm plus hb20

Table of Contents