Completing measurement
When the beam pointing
measurement is completed, the
indicator Beam Pointing changes to
green.
1.
Click on the Stop button.
2.
Switch off the laser.
You can reset the measured values in the
Position and Angle windows by pressing
the Reset button.
9.10

Fiber mode measurement

The measuring mode Fiber mode is - apart from one difference - identical to the measuring mode Manual
caustic (see Chapter "9.7 Manual caustic measurement" on page 73).
The difference is that the axes are locked in the x and y direction, i.e. they cannot be moved.
In this mode, it is shown below the green start button that the axes are locked.
This measuring mode is available only for the MSM+ HB10. With the MSM + HB20, the movement in the x
and y directions is always blocked.
Select this measuring mode instead of the Manual caustic measuring mode if you have installed a fiber
bridge. Without locking the axes, damage to the fiber or the fiber holder can occur.
88