Danger To The Device; Parameters Of Cooling Water Connection - Primes MicroSpotMonitor Plus HighBrilliance Original Instructions Manual

Table of Contents

Advertisement

MSM+ HB
Connecting cooling circuit at MSM+ HB20
There is no adapter required for the
X Connect the cooling water directly to the device.
8.9.2

Danger to the device

NOTICE
Damage/Destruction of the device
If the following requirements for the cooling circuit are not observed, the measuring device
may be damaged.
X Observe the following requirements.
Water quality
The device can be operated with tap water as well as demineralized water.
An operation with strongly deionized water (DI-water, conductivity < 30 µS/cm) is only possible with the res-
pective connection parts (stainless steel) – we would be glad to advise you if necessary.
Large dirt particles or teflon tape may block internal cooling circuits. Therefore, please thoroughly rinse the
system before connecting it.
Aluminum components
Do not operate the device on a cooling circuit in which aluminum components are installed. Otherwise, cor-
rosion in the cooling circuit can occur, particularly when operating at high powers and power densities. In the
long term, this will reduce the efficiency of the cooling circuit.
8.9.3

Parameters of cooling water connection

Data
PE hose diameter
Recommended flow rate
Minimum flow rate
Recommended water pressure
Maximum water pressure
Revision 00 EN - 02/2022
MSM+ HB20.
Water out
MSM+ HB10
12 mm
7 - 8 l/min
5 l/min
2 bar
4 bar
Water in
MSM+ HB20
16 mm
14 - 16 l/min
10 l/min
2 bar
4 bar
41