MSM+ HB
Option
Position in mm
Find beam
autom. Exposure
Integration time in ms
Filter level
autom. filter adjustment
Time series
Number of measurements *
Interval between measurements in s *
Parameter settings
Save current parameters
Load last parameters
Tab. 9.3:
Options in the Device control > Settings tab of the Single planes measuring mode
Revision 00 EN - 02/2022
Explanation
Use one of the following options to adjust the position of the
measurement window:
X Enter the x-position / y-position of the center of the measurement
window in the input field on the left / right.
X Position the mouse pointer within the measurement window. Then
drag while holding down the right mouse button.
This option enables a measurement in which the laser beam is auto-
matically searched for in the entire measurement range. The plane at
the defined z-position is then displayed in the graphic representation.
Measurement window size, integration time, and filter level of the filter
wheel are set automatically.
X Click on the button to start the beam search.
Note that the determined measurement data will not be saved in the
project tree of the Projects tab.
If this option is enabled, the integration time is set automatically.
X Set the check mark to enable the option.
If the autom. Exposure option is disabled, the integration time can be
set manually.
Use one of the following options:
X Enter a value in the input field (0.006 ms to 225 ms).
X Use the slider below the input field.
The selected setting will only be active if the autom. filter adjustment
option is disabled.
The filter required for measurement depends on the wavelength and
intensity of the laser beam to be measured. We recommend selecting
a filter level that achieves a detector modulation of between 60 % and
80%.
X Select a filter level in the drop-down list.
If this option is enabled, then the filter levels are set automatically.
X Set the check mark to enable the option.
A time series consists of several single plane measurements with the
same settings.
X Enter the desired number of single plane measurements of the time
series.
X Enter the pause between the single plane measurements of the time
series. This is the time gap between the end of one measurement
and the start of the next.
All settings in the Device control menu can be individually saved for
each device. The saving location is the local installation of the LaserDia-
gnosticsSoftware LDS.
X Click this button to save all current settings of the connected de-
vice.
X Click this button to load the last saved device configuration.