Safety Circuit - Primes MicroSpotMonitor Plus HighBrilliance Original Instructions Manual

Table of Contents

Advertisement

MSM+ HB
8.4

Safety Circuit

Connect this port to your laser system (Safety PLC or Emergency off).
The Safety Interlock protects the MSM+ HB from damage by switching off the laser in the following cases:
The voltage supply at the MSM+ HB is not connected or interrupted
There is an overtemperature at the absorber
A referencing procedure is triggered during the measurement
The interlock connection provides two redundant safety circuits.
NOTICE
Danger of damage due to overheating
If the safety circuit is not connected, the device can be damaged due to overheating.
X Only operate the device with a correctly connected safety circuit
X Check that the laser is switched off properly by the external safety circuit in the event of a
fault.
M8 connector 8-pin (Pin: view connector on device; Color: color of cable core)
2
1
8
3
7
4
6
5
Matching ready-made cable: Binder 79-3808-52-08
Tab. 8.2:
Pin assignment of the safety interlock connector
If the safety interlock is triggered, pin 1 and pin 2 (pin 4 and pin 5) are connected. If the values correspond to
the operating conditions, pin 1 and pin 3 (pin 4 and pin 6) are connected.
Pin 7 and 8 must be connected. Otherwise the interlock circuit is not active.
Revision 00 EN - 02/2022
Pin
Color
Function
1
White
COM 1:
2
Brown
NoClosed 1: Connected with pin 1 when not ready for operation
3
Green
NoOpen 1: Connected with pin 1 when ready for operation
4
Yellow
COM 2:
5
Grey
NoClosed 2: Connected with pin 4 when not ready for operation
6
Pink
NoOpen 2: Connected with pin 4 when ready for operation
7
Blue
+ 24 V, supply from external
8
Red
GND to external voltage (Pin 7)
Connected with pin 2 when not ready for operation
Connected with pin 3 when ready for operation
Connected with pin 5 when not ready for operation
Connected with pin 6 when ready for operation