Measure A Plane Continuously - Primes MicroSpotMonitor Plus HighBrilliance Original Instructions Manual

Table of Contents

Advertisement

MSM+ HB
9.6.4

Measure a plane continuously

1.
Switch on the laser.
2.
Click on the Settings tab.
3.
Click on the Find beam button.
 The laser beam is automati-
cally searched for in the entire
measurement range. Measurement
window size, integration time and filter
level of the filter wheel are set auto-
matically.
If the search is successful, the beam
is displayed in the graphic representa-
tion.
If the beam is not displayed:
4.
Select the Low power mode option
or increase the integration time or se-
lect a weaker filter of the filter wheel.
5.
Select a different z-position.
6.
Check again the correct alignment of
the device on the x-y plane.
7.
Increase the laser power (step by
step).
As soon as the beam is displayed:
8.
Click the Start button.
 The measurement begins.
Data is continuously read out from the
previously set measurement window
and displayed with a high frame rate
in the false-color image.
9.
To stop the measurement, click on
the Stop button.
10. Switch off the laser after the
measurement is completed, unless
you want to perform further measu-
rements.
Revision 00 EN - 02/2022