Quick overview installation

This quick overview informs you in advance about necessary protective measures, mediums necessary for
operation and required connecting elements.
1.
Take safety precautions
DANGER
Danger to life for persons with pacemaker or implanted defibrillator
Magnetic spring rotors consist mainly of neodymium magnets (NdFeB magnets). These
can impair the correct functioning of pacemakers.
X If you have a cardiac pacemaker or implanted defibrillator, keep a minimum distance of
1 m from the device.
2.
Aligne the device to the laser beam and mount stable
An alignment tool is included in the scope of delivery
You need 6 screws M8x1 and 2 screws for the mounting holes Ø 6.6 mm
3.
Install the water-cooling
Connection diameter 12 mm (16 mm for 20 kW-model)
Flow rate 7 l/min - 8 l/min (14 l/min - 16 l/min for 20 kW-model)
4.
Connect the compressed air
Compressed air according to ISO 8573-1:2010 [6:4:4].
0.5 bar - 1 bar
Connection diameter 6 mm
5.
Disassemble the inlet aperture cover
6.
Electrical connection
Establish voltage supply
External safety switch (interlock)
7.
Connect with the PC
Connection via Ethernet
8.
Install the LaserDiagnosticsSoftware LDS on the PC
Software is part of the scope of delivery
22
MSM+ HB
Chapter 1 on page 9
Chapter 7 on page 27
Chapter 8.9 on page 39
Chapter 8.10 on page 43
Chapter 6 on page 23
Chapter 8.2 on page 36
Chapter 8.3 on page 36
Chapter 9 on page 44
Revision 00 EN - 02/2022