Power Supply (24V); Ethernet (Lan) - Primes MicroSpotMonitor Plus HighBrilliance Original Instructions Manual

Table of Contents

Advertisement

8.2

Power supply (24V)

The device requires a supply voltage of 24 V ± 5% (DC) for operation. A suitable power supply unit with cable
is included.
The connection socket for power supply is a 5-pin, L-coded M12 connector.
Harting M12-P-PCB-THR-2PC-5P-LCOD-M-STR
3
4
FE
Tab. 8.1:
Connection socket for the power supply
8.3

Ethernet (LAN)

NOTICE
Damage / destruction of the device due to voltage peaks
Disconnecting the connecting cables during operation causes voltage peaks. This could
destroy the communication modules of the device.
X Connect / disconnect all plugs only when the power is off.
1. Connect the device with the PC or with the network.
2. Plug the power supply into the connector socket for the power supply.
Fig. 8.3:
Connection via Ethernet with a PC or a local network
36
Pole
1
2
2
3
4
1
5
PRIMES Power Supply
Ethernet
PC
Function
+24 V
Not connected
GND
Not connected
FE (functional earth)
or
Ethernet
LAN
MSM+ HB
Revision 00 EN - 02/2022