Configure Advanced Settings - Primes MicroSpotMonitor Plus HighBrilliance Original Instructions Manual

Table of Contents

Advertisement

Option
Filter level
autom. filter adjustment
Time series
Number of measurements *
Interval between measurements in s *
Tab. 9.9:
Options in the Device control > Settings tab of the Automatic caustic measuring mode
9.8.2

Configure advanced settings

1.
Click on the Advanced tab.
2.
Maintain the options according to the
explanations in the following table..
Option
Save device settings
# Pixel in x/y *
Number of planes *
Low power mode
Tab. 9.10:
Options in the Device control > Advanced tab of the Automatic caustic measuring mode
80
Explanation
The selected setting will only be active if the autom. filter adjustment
option is disabled.
The filter required for measurement depends on the wavelength and
intensity of the laser beam to be measured. We recommend selecting
a filter level that achieves a detector modulation of between 60 % and
80%.
X Select a filter level in the drop-down list.
If this option is enabled, then the filter levels are set automatically.
X Set the check mark to enable the option.
A time series consists of several caustic measurements with the same
settings.
X Enter the desired number of caustic measurements of the time
series.
X Enter the pause between the caustic measurements of the time
series. This is the time gap between the end of one measurement
and the start of the next.
Explanation
All options marked with an asterisk in the Device control menu can be
saved in the EEPROM of the device.
These and other options for saving / loading configurations are descri-
bed in chapter 9.4.4 on page 53.
This option enables setting the resolution of the CCD sensor:
X Enter the number of pixels in the x-direction / y-direction in the input
field on the left / right.
The default setting is 21 planes.
X Enter the number of planes to be measured.
We recommend enabling this option for relatively low power laser be-
ams.
If this option is enabled, the following initial values are selected during
beam search
Filter level of the filter wheel: 0 (instead of 1 otherwise)
Integration time: 10 ms (instead of 1 ms otherwise).
X Set the check mark to enable the option.
MSM+ HB
Revision 00 EN - 02/2022

Advertisement

Table of Contents
loading

This manual is also suitable for:

Msm plus hb10Msm plus hb20

Table of Contents