Option
Locked area
Maximum z-Position in mm *
Tab. 9.4:
Options in the Device control > Advanced tab of the Single planes measuring mode
62
Explanation
Locked area means a restriction of the traversing range in the z-direc-
tion. The lower limit of the enabled traversing range is always home
position. The upper limit can be set freely.
1.
Use one of the following options to set the locked area:
X Enter the upper limit of the permitted z-traversing range in the input
field.
X Use the slider below the input field.
We recommend saving a locked area in the device, if it is safety-relevant:
2.
If you want to save the locked area (and the other options marked
with an asterisk) in the device, click the Save device settings
button.
 The settings are retained even if the device is switched off or dis-
connected from the power source.
MSM+ HB
Revision 00 EN - 02/2022