Mounting The Fibre Bridge (Option) - Primes MicroSpotMonitor Plus HighBrilliance Original Instructions Manual

Table of Contents

Advertisement

1. Put the alignment tool on the cyclone. The upper surface corresponds to the z-position of the measuring
plane of the objective. The distance between lower reference plane and upper surface of alignment tool
is z.
2. Turn on the pilot laser. If the laser hits the small hole in the
on the sensor. Typically, the misalignment angle between the beam and the instrument axis should not
exceed a divergence of 10 mrad (0.5°).
3. If the laser beam does not hit the small hole vertically, you can align the MSM+ HB with the 6 screws in
the threaded holes.
4. Finally, fasten the device, so that it cannot be moved by accidental bumping or pulling on the cables.
Fig. 7.6:
Fastening bores MSM+ HB, view from above
2 mounting holes Ø 6,6 mm
6 thread holes M8x1 for the alignment
7.4

Mounting the Fibre Bridge (option)

To fix fibres - for example when using fibre lasers - a fibre bridge with fibre adapter can be mounted.
NOTICE
Damaging/destroying the fiber
In the upper z-axis position the device can collide with the plugged fiber and damage it
X Note the maximum travel range (chap. "14 Technical data" on page 97)
X Dismount the cyclone before the startup!
32
vertically, it is displayed centrally
alignment tool
Revision 00 EN - 02/2022
MSM+ HB