MSM+ HB
6.4

Draining the cooling circuit of the MSM+ HB

The cooling circuit of the MSM+ HB must be completely drained. Storage or transport at a temperature that
is close to or below freezing point can lead to damages due to the formation of ice, if the cooling circuit is not
completely empty.
X Drain the cooling circuit completely.
You may use compressed air to purge the cooling
circuit of the MSM+ HB.
The photo shows the situation at the MSM+ HB10
6.5

Sealing the cooling circuit of the MSM+ HB

Even if the cooling circuit has been drained completely, a small amount of residual water remains inside the
device. This may leak during transport and damage the electronics of the device.
X Seal the connecting plugs of the cooling circuit
by using of the enclosed plugs.
A total of 4 plugs with 12 mm outer diameter is
required for the MSM+ HB10.
A total of 2 plugs with 16 mm is required for the
MSM+ HB20
The photos show the situation at the MSM+ HB10
Revision 00 EN - 02/2022
Cooling circuit connectors
4
1
2
3
25