MSM+ HB
9.5.3
Measure a plane
DANGER
Serious eye or skin injury due to laser radiation
X Observe the warning messages in chapter 9.1 on page 44.
1.
Switch on the laser.
2.
Click on the Settings tab.
3.
Click on the Find beam button.
The laser beam is automati-
cally searched for in the entire
measurement range. Measurement
window size, integration time and filter
level of the filter wheel are set auto-
matically.
If the search is successful, the beam
is displayed in the graphic representa-
tion.
If the beam is not displayed:
4.
Select the Low power mode option
or increase the integration time or se-
lect a weaker filter of the filter wheel.
5.
Select a different z-position.
6.
Check again the correct alignment of
the device on the x-y plane.
7.
Increase the laser power (step by
step).
As soon as the beam is displayed:
8.
Click the Start button.
The measurement begins.
9.
Optional:
If necessary, click the Stop button to
end the measurement prematurely.
Revision 00 EN - 02/2022
63