Option
Use
Reset z-Axis position
Power P in W *
Filter level
autom. filter adjustment
Time series
Number of measurements *
Interval between measurements in s *
Tab. 9.7:
Options in the Device control > Settings tab of the Manual caustic measuring mode
74
Explanation
X Click this button to transfer the z-values displayed in the Caustic
suggestion fields to the z-Axis position (z1 / z2) in mm fields.
X Click this button to reset the z-axis position to the default values
To calculate the power density, the laser power used must be entered.
Otherwise, the measured amplitudes are given directly in counts.
X Enter the laser power used during measurement.
The selected setting will only be active if the autom. filter adjustment
option is disabled.
The filter required for measurement depends on the wavelength and
intensity of the laser beam to be measured. We recommend selecting
a filter level that achieves a detector modulation of between 60 % and
80%.
X Select a filter level in the drop-down list.
If this option is enabled, then the filter levels are set automatically.
X Set the check mark to enable the option.
A time series consists of several caustic measurements with the same
settings.
X Enter the desired number of caustic measurements of the time
series.
X Enter the pause between the caustic measurements of the time
series. This is the time gap between the end of one measurement
and the start of the next.
MSM+ HB
Revision 00 EN - 02/2022