Compressed Air

The compressed air is required in order to generate an air flow in the cyclone which is directed outwards.
This is supposed to prevent the penetration of dirt particles.
NOTICE
Damage to the optical components
Contaminated compressed air can permanently damage the optical components of the mea-
suring device.
X The compressed air must be according to specification ISO 8573-1: 2010 [6: 4: 4].
Connect the compressed air supply by means of a plastic hose.
Data
Hose diameter
Typical air pressure
Maximum permissible air pressure
Specification according to
The compressed air connector is located opposite the connection side of the device.
Compressed air
supply ∅ 6 mm
Fig. 8.6:
Compressed air supply
CAUTION
Hearing damage
If the compressed air hose is loosened when pressure is still applied to the hose, this gene-
rates noise. This can cause hearing damage.
X Only loosen the compressed air hose when there is no longer any pressure.
Connecting the compressed air hose
X Push the hose into the plug-in connection as far as it will go.
Disconnecting the compressed air hose (Only when the measuring objective is closed)
X Press down the blue release ring of the connector with one hand and pull out the hose with the other
hand.
Revision 00 EN - 02/2022
Value
6 mm
0.5 - 1.0 bar
2 bar
ISO 8573-1: 2010 [6: 4: 4].
43