6.6
Sealing the aperture of the HB objective
The magnetically held alignment tool may fall off during shipping. This can lead to contaminations within
the objective. The objective of the MSM+ HB must therefore be sealed for shipping with the enclosed black
aperture lock.
1. Turn on the compressed air.
2. Remove the alignment tool.
3. Disassemble the cyclone.
4. Install the aperture lock and the sealing ring with
4 cylinder head bolts M3x8.
5. Turn off the compressed air.
6.7
Packing the device
1. Pack the device completely in a plastic wrap to avoid contamination with particles from the case padding
foam.
2. Place the packaged unit in the PRIMES transport case and close the cover with all tension locks. All ac-
cessories must be completely enclosed.
The fiber bridge belonging to the device (option) must be enclosed with each service shipment. Else a
correct alignment of the bridge to the device cannot be guaranteed performing mechanical work on the
MSM+ HB.
1. Uninstall the fiber bridge.
2. Pack the fiber bridge separately.
26
MSM+ HB
Aperture lock
Sealing ing
Fiber bridge
Revision 00 EN - 02/2022