Aligne the MSM+ HB with the Monitor measuring mode

Using a pilot laser:
You can use the continuous plane measurement to align the MSM+ HB. Use the pilot laser for this purpose.
1.
Switch on the pilot laser.
2.
Click on the Advanced tab.
3.
Activate Low power mode
4.
Click on the Settings tab.
5.
Click on the Find beam button.
 The laser beam is automatically
searched for in the entire measuring
range.
After a successful search, the beam is
displayed in the graphical representa-
tion..
Using a laser of laser class 4:
DANGER
Serious eye or skin injury due to laser radiation
X Only align the device using a laser (laser class 4) if the process is carried out remotely
behind a separating protective equipment. The protective equipment must be suitable for
blocking the radiation or attenuating it to a non-hazardous level. Use the laser's low power
mode.
Only align the device by means of a laser of laser class 4 if
alignment with the pilot laser is not possible and
the process is carried out remotely behind a separating protective equipment. The protective equipment
must be suitable for blocking the radiation or attenuating it to a non-hazardous level. Use the lowest
possible power on the laser.