Connect / disconnect the device with the LaserDiagnosticsSoftware LDS

9.3.1

Switch on the device and connect it to the LDS

CAUTION
Crushing hazard
The camera housing of the device can be moved along the z-axis.
X Never reach into the traversing range of the camera housing.
NOTICE
Damaging / destroying the device
The camera housing of the device can be moved along the z-axis. Obstacles in the traver-
sing range of the device can lead to collisions and damage the device.
X Always keep the traversing range of the camera housing free of obstacles.
If the camera housing was not in the home position when the device was switched off / disconnec-
ted from power source, then it moves to the home position after it is switched on / connected to
power source. This happens even before the device is connected to the LaserDiagnosticsSoftware
LDS. The axes are referenced to the home position.
Observe the correct sequence:
All electrical connections must be established before the device is switched on (see Kapitel "8
Connections" on page 35).
The device must be switched on before the LaserDiagnosticsSoftware LDS is started.
1.
Switch on the device.
The operating status is shown in the
status display. If the device is ready
for operation, the display lights up
white.
If necessary, the camera housing
moves to the home position.
2.
Use one the following options to start
the LaserDiagnosticsSoftware LDS:
X Double-click on the program icon in
the start menu group.
X Double-click on the desktop shortcut.
The start logo appears for a short
time.
Revision 00 EN - 02/2022
On/Off switch
Status display
47