General procedure for measuring

This chapter contains general information about the LaserDiagnosticsSoftware LDS. Read this general infor-
mation before turning to the following chapters on the different measuring tasks.
9.4.1
Open the „Device control" menu
After connecting the device, the Device control menu must be opened:
1.
Click on the Devices tab.
2.
Select the device and click on the de-
vice function CCD below the device
name.
The Device control menu with the
Adjustment and Measuring modes
sections opens.
9.4.2

Move camera housing to desired z-position

The Adjustment area of the Device control menu offers various options for moving the camera housing.
Option
z-Position in mm
Move to axis position
Move to axis center position
Tab. 9.2:
Options in the Adjustment area of the Device control menu
Revision 00 EN - 02/2022
Device function
Explanation
X Enter the desired z-position in the input field or use the slider.
X Click this button to move the camera housing to the z position spe-
cified in the z-Position in mm field.
X Click this button to move the camera housing to the center position
of the (unlocked) z-axis traversing range.
Note that this does not have to be the maximum traversing range of
your device. If a locked area of the z-axis was stored in the EEPROM
during a previous use of the device, then this locked area is active. The
camera housing then moves to the center position of the unblocked
area.
51