SEMI-AUTOMATIC CEILING SUSPENSION WITH RAD WALL STAND DR
Both the DR Detector and the Ceiling Suspension can be Master and Slave, but
the Auto-tracking movement is always in vertical direction.
Proceed as indicated for manually operated Auto-tracking:
1.
Press the AUTO-TRACKING control of the Console (or Remote Control
if available) to activate the function, the background is in yellow when
activated.
2.
Move manually the Master equipment (refer to Illustration 4-26):
a.
The RAD Wall Stand DR is the Master and is moved vertically. The
DR Detector can be tilted at any angle or in vertical position at 0
Press and hold the Vertical Lock or step on the UP/DOWN Pedal
of the Footswitch. The Ceiling Suspension tracks its movement in
the same direction in all cases. The SID remains constant.
b.
The Ceiling Suspension is the Master and is moved vertically. The
DR Detector must be at 0
button of the Control Console and move up/down the X-ray Tube.
The DR Detector moves in the same direction. The SID remains
constant. It is not valid when the DR Detector is tilted at 90
3.
Keep pressing the brake or pedal until the Slave equipment arrives to the
final position and is aligned with the Master Equipment.
4.
In case the brake button is released before finishing the Auto-tracking
movement, it gets aborted. Once the brake button or pedal is pressed
again, the Slave equipment gets aligned with the Master equipment at
the default SID.
o.
Press and hold the Vertical Brake