6.5
ALIGNMENT
Illustration 6-10
DR Detector Center Mark
0242F EN 20181218
The three field pattern on the Front Panel of the RAD Wall Stand corresponds
to the three detection areas for the Ion Chamber Detector.
It is important that the X-ray Tube is accurately centered with the DR Detector
transversely. If the alignment is not accurate, density cut-off at the edges of the
image and appearance of grid patterns may be found.
To assure that the DR Detector is aligned with the X-ray beam, move the
DR Detector or the X-ray Tube, in order that the collimator light is aligned with
this center mark of the Front Panel.
DX-D 600
User Manual
127