GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
The DX-D 600 Radiographic Room is intended for use in the electromagnetic environment specified below.
The customer or Operator of the DX-D 600 Radiographic Room should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines.
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
0242F EN 20181218
(IEC 60601-1-2:2007)
IEC 60601-1-2:2007
Test Level
¦ 6 kV contact
¦ 8 kV air
2 kV for power supply lines
¦
1 kV for input/output lines
¦
1 kV line(s) to line(s)
¦
2 kV line(s) to earth
¦
<5 % UT
( >95 % dip in UT)
for 0.5 cycles
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
NOTE - U
is the a.c. mains voltage prior to application of the test level.
T
Compliance Level
6 kV contact
¦
8 kV Air
¦
2 kV for power supply lines
¦
1 kV for input/output lines
¦
1 kV symmetrical coupling
¦
2 kV asymmetrical coupling
¦
>95 %
for 0.5 periods
60 %
for 5 periods
30 %
for 25 periods
>95 %
for 250 periods
3 A/m
DX-D 600
User Manual
Electromagnetic
Environment - Guidance
Floors should be wood, concrete or
ceramic tile.
If floors are covered with synthetic
material, the relative humidity
should be
at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
DX-D 600 Radiographic Room
requires continued operation during
power mains interruptions, it is
recommended that the DX-D 600
Radiographic Room is powered
from an Uninterruptible Power
Supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
39