4.2.1
CEILING SUSPENSION MANUAL MOTION
.
Note
0242F EN 20181218
BEFORE POWERING ON AND MOVE THE UNIT, CHECK
THAT THERE IS NO OBJECT OR OBSTACLE ON THE TUBE
SUPPORT OR THE L-BLOCK SURFACE FOR THE CORRECT
MOTION OF THE SUSPENSION.
THIS EQUIPMENT CAN BE MOVED IN DIFFERENT AXES.
PLEASE TAKE CARE THAT NEITHER THE PATIENT NOR
OPERATOR/STAFF ARE IN THE MOVEMENT AREA OF THE
EQUIPMENT. ALWAYS WATCH WHERE YOU STANDING.
REMOVE ALL OBJECTS FROM THE COLLISION AREA.
IT IS MANDATORY TO POSITION FIRST THE EQUIPMENT AT
THE INITIAL POSITION OF THE RAD EXAMINATION AND
THEN WITH THE SYSTEM ALREADY STOPPED, POSITION
THE PATIENT.
THIS EQUIPMENT MUST BE CONTROLLED JUST BY ONE
OPERATOR. SIMULTANEITY OF ACTIONS IN MOVEMENT
OR RADIATION PUSHBUTTONS MUST BE AVOIDED. ONLY
IF AN SPECIAL ACTION IS INDICATED IN THIS MANUAL
SIMULTANEITY OF ACTIONS IS ALLOWED.
IN THE EVENT OF AN EMERGENCY, TURN OFF THE CEILING
SUSPENSION PRESSING FORCIBLY THE EMERGENCY OFF
SWITCH (RED MUSHROOM SHAPED SWITCH) ON THE
X-RAY
TUBE
CONTROL BOX OR AT THE ROOM ELECTRICAL CABINET.
To move the equipment in relation with its axes:
1.
Press and hold the corresponding button of the axis Brake on the Control
Console. The Brake will be released.
2.
Drive the Ceiling Suspension to the desired position.
3.
Release the button. This will re-energize the brake.
To carry out freely all movements simultaneously on the Vertical, Transverse
or Longitudinal Axes, use the Omni-directional Button located on the Wheel of
the Ceiling Suspension.
The equipment stops moving at Detent Points even when the
brake has been released. Press and hold the button again to
continue with the movement.
SUPPORT,
AUTOMATIC
DX-D 600
User Manual
POSITIONING
69

Hide quick links: