Illustration 4-17
Patient Virtual Safety Areas
0242F EN 20181218
Volumetric protection areas are activated around both DR Detectors. A
virtual safety area has to be configured around a patient lied down on the
RAD Table or positioned at the RAD Wall Stand DR.
The equipment speed will be limited to one-half and an audible signal will
be activated when the equipment is within 305 mm (12") of the patient
Tabletop or within 100 mm (4") of the RAD Table side.
Manual movements are free inside these protection areas, but automatic
movements are forbidden from outside to any place inside this area.
Automatic movement within the virtual safety area is allowed to
escape from this area, but at a reduced speed.
In these areas, the motion will not continue more than 13 mm (0.5") after
the Operator has activated the stop or motion reverse control.
Minimum SID for automatic movements >950 mm (37.4").
Maximum SID for automatic movements <4000 mm (137.5"), if Room
configuration and rails dimensions allow it.
In case of collision the pressure is limited to 178.6 kg/m (10 lbs/inch)
maximum and 178 Newtons (40 lbs.)
If the Operator changes the DR Detector position during the
Auto-centering movement, the Ceiling Suspension stops this movement
and goes back to the initial position.
It will not be possible to complete any automatic movement when the
Input Interlock Signal is active.
305 mm (12")
DX-D 600
User Manual
100 mm (4")
79