AGFA DX-D 600 User Manual page 129

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

6.1
RAD WALL STAND DR MOVEMENTS
6.1.1
VERTICAL MOVEMENT
Illustration 6-2
Press Vertical Lock Handle to release the DR Detector Brake
DR Detector Brake
.
Note
0242F EN 20181218
The RAD Wall Stand DR enables radiographic operations at different positions
within the range of the Vertical Carriage travel, adjustable tilting angles as well
as rotation of the DR Detector.
MONITOR WITH SPECIAL CARE THE PATIENT POSITION,
HANDS, FEET, FINGERS, ETC.) AND USE THE PATIENT
HAND-GRIPS TO AVOID INJURY TO PATIENT CAUSED BY
UNIT MOVEMENTS. PATIENT HANDS MUST BE KEPT AWAY
FROM MOBILE COMPONENTS OF THE UNIT.
The DR Detector assembly can be moved manually by pressing the Brake
Control on the Vertical Handle. The DR Detector assembly can be moved
automatically when executing any Auto-position or Programmed Position that
has been configured with a DR Detector vertical displacement.
1.
Press the Brake Control on the Vertical Lock Handle, it is indicated by a
label near the handle.
2.
Check that the brakes are released and the DR Detector can be moved
smoothly up and down.
3.
Set the DR Detector assembly at the desired height, depending on the
study to be performed.
4.
Release the Brake Control on the Vertical Lock Handle, the Brake is
activated and the DR Detector is locked.
The Vertical Lock Handle is factory located at the left side, behind
the DR Detector. If needed, it can be field configured at the right
side by the Service Enginner.
DX-D 600
User Manual
121