0242F EN 20181218
There are different kind of Detents Points:
Floating and Electronic Detent Points. Both systems are provided
with floating Detent Points by default. These points are defined by
software in every final position of a PP or Auto-position:
In case of the Automatic System refer to Section 4.2.4 - "Position.
G
Automatic System Auto-positioning". and Section 4.2.6 -
"Programmed Positions (PP)".
In case of the Semi-automatic System refer to Section 4.2.5 -
G
"Position. Semi-automatic System Auto-positioning" and Section
4.2.6 - "Programmed Positions (PP)", electronic detents are
optional.
Software defined Detent Points during manual motion. In some
cases it is possible to configure up to 4 different detent positions during
for each axis and workstation. This option is available just if
configuration is changed by customer technical service. At these
Detent Points, the equipment will activate the brakes, whenever the
manual movement is made smoothly, when the movement is faster than
configured then Detent Point is skipped.
Mechanical Detents for Semi-automatic Ceiling Suspension. The
Mechanical Detents kit are provided for the Longitudinal and Transverse
Axes. Install these mechanical detent kit when it is required to improve
the Auto-tracking detent points performance. It is not mandatory their
installation when the electronic detents are enabled.
There are some audible signals that alert when:
the equipment enters/exits into the Detent Point influence area.
the Detent Point is activated.
To avoid that the Ceiling Suspension stops at these Detent Points, just move
faster the Ceiling Suspension to avoid the detent points or press on the Detent
Skipper Button to deactivate all Detent Points.
By default there is one Detent Point configured in the center of both Receptor
(Table and Wall Stand). If the Auto-tracking is activated this Detent is
auto-configured in this initial point, if the Receptor is moved to a new position
and the Auto-tracking is activated again, the new position will be configured
automatically as the new Detent Point.
DX-D 600
User Manual
71