4.2.6
PROGRAMMED POSITIONS (PP)
.
Note
.
Note
.
Note
Illustration 4-30
Programmed Positions Area Display
PP Description
Validate PP
98
Each PP refers to a position of the X-ray Tube (SID, ANG & ROT), but also to
the DR Detector of the RAD Table on vertical and longitudinal travels, and to
the DR Detector of the RAD Wall Stand on vertical travel and on tilting angle.
It can be configured also to have the Auto-tracking movement active.
With the Programmed Positions, the Tube position is referenced to a
Workstation configured also in the PP.
In the Semi-automatic System, this function allows to configure
automatic movements in the Vertical Axis only. It is important that
the RAD Wall Stand and RAD Table were configured correctly
before executing the Programmed Position.
In case the last PP was configured with the Auto-tracking On, it
remains ON until it is deactivated, manually or selecting other PP
configured with the Auto-tracking OFF.
All PP are previously configured in the field by the Service
Engineer, they can not be modified directly by the operator. To
modify them, contact Service.
PP ID #