User Manual
Table 8-9
Auto-tracking
CODE
USER ERROR MESSAGE
Tracking aborted: No Detector
11903
Operative rejected:
11906
Detector and Tube-crane are
misaligned: Please, get a valid SID
System geometry: Positioner out of
11907
Auto-tracking aborted:
11913
Detector changed
11914
Auto-tracking aborted: CS moved
Auto-tracking aborted:
11915
Wall Stand moved
11916
Auto-tracking aborted: Table moved
Auto-tracking aborted:
11921
CS-Longitudinal limit
Auto-tracking paused:
11922
CS-Longitudinal limit
Auto-tracking aborted:
11931
CS-Transversal limit
Auto-tracking paused:
11932
CS-Transversal limit
Auto-tracking aborted:
11941
CS-Vertical limit
Auto-tracking paused:
11942
CS-Vertical limit
Auto-tracking aborted:
11981
Wall Stand-Vertical limit
Auto-tracking aborted:
11991
Table-Longitudinal limit
Auto-tracking paused:
11992
Table-Longitudinal limit
158
selected
Bucky/Detector and X-ray Tube are
limits
Ceiling Suspension moved manually
Table moved manually and misaligned
Target position out of calibrated limits
on Longitudinal Axis of the Ceiling
Target position out of calibrated limits
on Transversal Axis of the Ceiling
Target position out of calibrated limits
Target position out of limits on Vertical
Target position out of limits on Vertical
Target position out of calibrated limits
ERROR DESCRIPTION
No Bucky/Detector selected
misaligned
Positioner out of limits
Target Bucky/Detector changed
position
and misaligned
Wall Stand moved manually and
misaligned
Suspension
Target position out of limits on
Longitudinal Axis of the Ceiling
Suspension
Suspension
Target position out of limits on
Transversal Axis of the Ceiling
suspension
on Vertical Axis of the Ceiling
Suspension
Axis of the Ceiling suspension
Axis of the Wall Stand
on Longitudinal Axis of the Table
Target position out of limits on
Longitudinal Axis of the Table
ACTION
Select Bucky/Detector and try again
Please get the X-ray Tube aligned
Please Place SID: between 80 and
300 cm
Try again
Activate Tracking function
Activate Tracking function
Activate Tracking function
Calibrate the Ceiling Suspension
longitudinal Axis
Do not exceed the longitudinal limit
Calibrate the Ceiling Suspension
transversal Axis
Do not exceed the transversal limit
Calibrate the Ceiling Suspension
vertical Axis
Do not exceed the vertical limit
Calibrate the Ceiling Suspension
vertical Axis
Calibrate the Table
Do not exceed the Table limits
0242F EN 20181218