GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC EMISSIONS
The DX-D 600 Radiographic Room is intended for use in the electromagnetic environment specified below.
The customer or the user of the DX-D 600 Radiographic Room should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
NOTE - In accordance with Standard IEC 60601-1-2:2014, the emissions characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 Class A. If it is used in a residential environment (for which CISPR 11 Class B is normally required) this equipment might not
offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or
re-orientating the equipment.
38
It is customer responsibility to assure that this equipment
and vicinity equipment complies the value of radio
frequency interferences shown in General Regulation for
safety according to IEC 60601-1-2:2007 and 2014 Tables as
described in this section.
The manufacturer is not responsible for any interference
caused by using other than recommended interconnect
cables, accessories and transducers or by unauthorized
changes or modifications to this equipment.
(IEC 60601-1-2:2007 AND IEC 60601-1-2:2014)
Compliance
Group 1
Class A
Class A
Complies
Electromagnetic environment - guidance
The DX-D 600 Radiographic Room use RF
energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The DX-D 600 Radiographic Room is suitable for
The DX D 600 Radiographic Room is suitable for
use in all establishments other than domestic and
those directly connected to the public low-voltage
power supply network that supplies buildings
power supply network that supplies buildings
used for domestic purposes.