4.2.4
POSITION. AUTOMATIC SYSTEM AUTO-POSITIONING
.
Note
.
Note
Illustration 4-27
Position Display Option
94
Configure directly in the Position screen of the Control Console the X-ray Tube
position manually or using any Auto-position. Up to 16 Auto-position can be
programmed in the field, each one refers to configured parking positions of the
X-ray Tube (SID, ANG & ROT) and of the DR Detectors (RAD Table
DR Detector on the longitudinal axis and RAD Wall Stand DR Detector on the
vertical axis). With the Auto-position, the Tube position is not referenced to any
specific Workstation, so all or just some of the available axes of the X-ray Tube
and DR Detectors can be configured on each Auto-position.
Auto-positions are intended to be used as parking positions or to
work with Direct Workstation. For a complete positioning of the
Ceiling Suspension, use the Programmed Positions.
The Auto-positions are configured by the Service engineer and
cannot be modified directly by the user. To modify Auto-positions,
contact Service.
Controls for Manual Position
Validate Manual Position
Auto-position Controls
Validate Auto-position