AGFA DX-D 600 User Manual page 103

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

0242F EN 20181218
1.
MANUAL POSITION
Select the Workstation.
Use the INCREASE/DECREASE Buttons to modify the Position
Parameters. Which are:
SID: Source-Image Distance from 800 to 2000 mm (31.5" to
G
78.7").
ANG: X-ray Tube angle degrees on Alpha Axis from -90 to 90
G
ROT: X-ray Tube rotation degrees on Beta Axis from -90 to 90
G
Press the VALIDATE MANUAL POSITION button to activate the
position.
Press and hold the PARKING Button on the Automatic Positioning
Control Box to execute the movement.
2.
AUTO-POSITIONS
Select the Workstation.
Select the Auto-position with the INCREASE/DECREASE Buttons.
Press the VALIDATE AUTO-POSITION Button to activate the position.
Press and hold the PARKING Button on the Automatic Positioning
Control Box to execute the movement.
Once the new position is reached, the PARKING POSITION Indicator lights up.
DUE TO SAFETY REASONS, WHEN A NEW AUTO-POSITION
IS SELECTED, IF THE FOCAL SPOT (TUBE) DISPLACEMENT
FROM THE ORIGINAL POSITION TO THE FINAL POSITION IS
>300 mm (11.8"), THE TELESCOPIC COLUMN WILL BE
RETRACTED TO A TRAVEL POSITION TO CARRY ON THE
MOVEMENT AND FINALLY ACHIEVE THE REQUIRED
POSITION. IF THE DISPLACEMENT DISTANCE IS <300 mm
(11.8") THE CEILING SUSPENSION WILL BE MOVED
DIRECTLY.
DX-D 600
User Manual
o
.
o
.
95