AUTOMATIC CEILING SUSPENSION WITH RAD TABLE
Depending on the Axis:
Horizontal Auto-tracking. Both pieces of equipment can
G
operate as Master or Slave.
Vertical Auto-tracking. The RAD Table is always the Master
G
element and the Ceiling Suspension the Slave.
Proceed as indicated for manually operated Auto-tracking:
1.
Press the AUTO-TRACKING control of the Console (or Remote Control
if available) to activate the function. The background is in yellow when
activated.
2.
Move manually the Master equipment (refer to Illustration 4-23):
a.
The Ceiling Suspension is the Master and is moved longitudinally.
Press and hold the Longitudinal brake Button of the Control
Console and move to the right/left. The DR Detector tracks its
movement in the same direction. The SID remains constant.
b.
The RAD Table is the Master and the DR Detector is moved
longitudinally. Press and hold the DR Detector brake button and
move to the right/left. The Ceiling Suspension tracks its movement
in the same direction. The SID remains constant.
c.
The Ceiling Suspension is the Master and is moved along its Alpha
Axis. Press and hold the Alpha Brake button of the Control
Console and angle the X-ray Tube. The DR Detector tracks its
movement longitudinally. The SID is modified.
d.
The RAD Table is the Master and is moved longitudinally while the
Ceiling Suspension is angled. Press and hold the DR Detector
brake button and move to the right/left. The Ceiling Suspension
tracks the movement of the DR Detector and moves longitudinally.
The SID remains constant.
e.
The RAD Table is the Master and is moved in the Vertical Axis.
Step and hold the RAISE or LOWER Control Pedal to move
UP/DOWN the DR Detector. The Ceiling Suspension tracks
vertically its movement in all cases. The SID remains constant.
3.
Press/Step the brake or pedal and hold until the Slave equipment arrives
to the final position and is aligned with the Master Equipment.
4.
In case the brake button is released before finishing the Auto-tracking
movement, it gets aborted. Once the brake button or pedal is pressed
again, the Slave equipment gets aligned with the Master equipment at
the default SID.