AGFA DX-D 600 User Manual page 53

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

2.9
QUANTITATIVE INFORMATION
.
Note
2.9.1
FUNCTIONAL TESTS PERFORMED TO OBTAIN THE QUANTITATIVE INFORMATION
0242F EN 20181218
The following tables show the Quantitative Information associated
to this X-ray System according with the Standard IEC
60601- -1- -3:2008 and IEC 60601- -1- -3:2008/AMD1:2013. These
tables illustrate loading factors for image performance and supply
Dose indication examples. Therefore, they are an example of the
adjustment of Loading Factors, Focal Spot Selection, SID and
Collimator opening, which affect to the radiation quality or to the
radiation dose rate applied in normal use.
Equipment:
Rad Positioner with Ralco Collimator.
Instrumentation used:
Dosimeter: Vacudap
Dosimeter: Unfors
Rectangular Phantom made of Polymethyl-methacrylate (PMMA)
layers: 25 cm x 25 cm x 20 cm.
Test Details:
Minimum SID distance from Table: 100 cm.
Maximum SID distance from Wall Stand: 180 cm.
Open Collimator size: 13 cm x 13 cm (min.), 43 cm x 43 cm (max.)
The measurements were made with the exposure parameters shown on
the results table:
KVp Range: 40 KVp, 60 KVp, 80 KVp, 100 KVp, 125 KVp
mAs Range: 1 mAs, 2 mAs, 10 mAs, 50 mAs, 100 mAs
Performed measurements of Air Kerma or Air Kerma Rate at the
following designated positions:
-- Distance SID doses
-- Patient (Phantom) Entrance doses and Entrance doses Rate
-- Patient (Phantom) Output doses and Output doses Rate
-- Collimator Output doses
DX-D 600
User Manual
45