AGFA DX-D 600 User Manual page 119

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

5.4
PORTABLE DR DETECTOR LOADING
Illustration 5-4
DR Detector Loading
5.5
GRID LOADING AND REMOVAL
0242F EN 20181218
1.
Using the DR Detector Handle, pull out the Tray to insert the DR Detector.
Press and hold the Handle brake until the Tray is completely pulled out.
2.
Insert the Detector to the back of the Tray. Push slightly the back
end-stops with the Detector for a correct fixation and fit it with the front
end-stops.
1
3.
Push in the tray using the Handle.
Check that the DR Detector is correctly inserted. A click sound means that the
Grid is in place.
Grids are intended to reduce scattered radiation and significantly enhance
image quality.
The RAD Table holds a removable grid focalized and labelled 100 cm (40"). The
Grid is provided with a label indicating its features.
Follow the procedure below for Grid loading. For Grid removal, follow the
procedure below in reverse order.
Handle Grid with care and place it in the accessories holder
when not in use. Dropping the Grid could cause damage and
reduced image quality.
2
DX-D 600
User Manual
111