8.4.2
ERROR LIST
Table 8-1
System/Miscellaneous
CODE
USER ERROR MESSAGE
9
Collimator cause
10
Interlock cause
11
Collimator Mode
12
Collimator Error
13
Stitching Cause
15
Dosimeter not communicating
16
Dosimeter not communicating
17
Dosimeter not communicating
18
Dosimeter not communicating
19
Collimator not communicating
0242F EN 20181218
Following tables list all defined error codes:
Error Code.
User Error Message. Displayed error message.
Error Description: Explains the error and reasons.
Action: Suggests the action to be completed in order to correct the error.
These errors are divided into nine different categories:
1
System/Miscellaneous
2
Ceiling Suspension Motor Controller Boards
3
RAD Wall Stand Motor Controller Boards
4
RAD Table Motor Controller Boards
5
Positioning
6
Safety Messages and Tomography
7
Auto-position ing
8
Auto-centering
9
Auto-tracking
ERROR DESCRIPTION
DAP NOT communicating: Test
DAP NOT communicating: Initialization
DAP NOT communicating: Heartbeat
DAP NOT communicating: Measure
Collimator NOT communicating
DX-D 600
User Manual
0 to 11099
11100 to 11199
11200 to 11299
11300 to 11399
11500 to 11599
11600 to 11699
11700 to 11799
11800 to 11899
11900 to 11999
ACTION
Click on Collimator Information
to know the reason
Check if Interlock is activated. Click on
Interlock icon to know the reason
Click on Collimator Information
to know the reason
Check SID/aperture relation
Try again
Contact Service
Contact Service
Contact Service
Contact Service
Contact Service
143