AGFA DX-D 600 User Manual page 120

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

DX-D 600
User Manual
Illustration 5-5
RAD Table Grid
Removable Grid
Grid Handle
SID Label
5.6
ACCESSORIES
5.6.1
HAND GRIPS
.
Note
112
1.
Pull out the Grid Tray.
2.
Install the Grid with the tube side label facing up.
3.
Push in the Grid Tray.
Check that the Grid is correctly inserted in the slot. A click sound means that
the Grid is in place.
The two Hand Grips are used to get patient hands away from the Tabletop
edges and make the patient feel secure when it is moving. They do not support
patients weight, but give patients a feeling of security and to avoid injuries.
For safety reasons, Hand grips must be used in all radiological examinations.
They can be installed along the Tabletop rails, fit them to the rails and locked
at any position with the thumbscrews.
USE ALWAYS THE HAND GRIPS TO AVOID INJURIES IN
PATIENT HANDS OR FINGERS WHEN THE TABLETOP IS IN
MOVEMENT. PATIENT'S HANDS MUST BE KEPT FAR AWAY
FROM THE TABLETOP EDGES IN EVERY MOMENT.
The Hand grips must not be positioned in the trajectory of the X-ray
beam.
Grid Slot
0242F EN 20181218