7.1.1
AUTOMATIC MODE
.
Note
7.1.2
BUSY MODE
MESSAGE
FOV
STS
USER
0242F EN 20181218
The Automatic mode is always activated whenever all the Positive Beam
Limitation (PBL) conditions are complied with:
The Aperture capacity must be enough to get a Field of View (FOV)
according to the std. IEC60601-1-3.
The Angle of the X-ray Beam must be orthogonal to the DR Detector, the
tolerance range is
Collimator position must correspond to 0
The X-ray Tube and DR Detector must not be rotated, at 0
The Ceiling Suspension must be in PREP status and pointing to the DR
Detector.
If any of this conditions is not complied with, the collimator automatically is in
Manual or Busy mode.
After selecting the Automatic mode from the Manual Mode, check
if it is necessary to change the FOV. It already remains as
configured for the Manual Mode.
This mode activates the X-RAY INTERLOCK, so it is not possible to do any
exposure. In Status Area appears the Interlock Icon and the description of the
reason of the Interlock. The reasons of the Busy Mode:
The blades aperture is being changed automatically.
It refers to the Collimator Busy Mode. When there is a new demand.
The blades aperture is being changed manually, using the manual Blades Controls.
o
3
.
¦
o
DESCRIPTION
DX-D 600
User Manual
of rotation of the X-ray Tube.
o
.
137