Table 8-2 (Cont.)
Ceiling Suspension Motor Controller Boards
CODE
USER ERROR MESSAGE
11121
Board Failure: CS-Angular
11122
End-Switch Limit: CS-Angular
Operative rejected: CS-Rotational
position out of calibrated range:
11123
Right there, only MANUAL
movement allowed
11124
Board Failure: CS-Rotational
11125
End-Switch Limit: CS-Rotational
11126
Control Board Error: CS-Interface
0242F EN 20181218
ERROR DESCRIPTION
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
DX-D 600
User Manual
ACTION
Change board
Check End Switch Limit position,
activation, operation.
Check End Switch Limit range.
Check position feedback.
Contact Service
Change board
Check End Switch Limit position,
activation, operation
Check End Switch Limit range
Change board
147