Table 8-6
Safety Messages & Tomography
CODE
USER ERROR MESSAGE
Tomography: Auto-positioning
11601
movement timeout: Please, check
positioning parameters
Tomography aborted: The Positioner is
11602
out of the allowed tolerance: Please,
try again
11603
Tomography: Illegal rotation axis
Tomography: The Generator interface
11604
board is not synchronizing: Please,
check the equipment state
Tomography: The Table is too high to
11605
get the SID: Please, lower the
equipment to for CS Auto-positioning
Tomography: Action has been aborted
11606
by user: Please, try again
Tomography: The cut plane must be
11607
refreshed
Tomography: The CS-Longitudinal
11608
(pre-tomo) target out of specifications:
Please, adjust site parameters
Safety Issue:
11610
Abnormal Velocity detected on
CS-Longitudinal
Safety Issue:
11611
Abnormal Velocity detected on
CS-Transversal
Safety Issue:
11612
Abnormal Velocity detected on
CS-Vertical
Safety Issue:
11613
Abnormal Velocity detected on
CS-Angular
Safety Issue:
11614
Abnormal Velocity detected on
CS-Rotation
Safety Issue:
11615
Abnormal Velocity detected on
Table-Longitudinal
Safety Issue:
11616
Abnormal Velocity detected on Wall
Stand-Vertical
0242F EN 20181218
ERROR DESCRIPTION
Auto-positioning movement timeout.
The Positioner is out of the allowed
tolerance
Rotation angle not allowed
The Generator interface board is not
synchronizing
The Table is too high to get the SID
Action has been aborted by user
The cut plane must be refreshed
The CS-Longitudinal (pre-tomo) target
out of specifications
Velocity upper-limit detected on
Suspension Longitudinal Axis
Velocity upper-limit detected on
Suspension Transversal Axis
Velocity upper-limit detected on
Suspension Vertical Axis
Velocity upper-limit detected on
Suspension Angular Axis
Velocity upper-limit detected on
Suspension Rotation Axis
Velocity upper-limit detected on Table
Longitudinal Axis
Velocity upper-limit detected on Wall
Stand Vertical Axis
DX-D 600
User Manual
ACTION
Please, check positioning parameters
Calibrate Table Receptor movement
Correct Rotation value and try again
Check configuration
Place the table height for enable Tomo
SID
Try again
Change cut plane
Please, adjust site parameters
In Manual motion, move smoothly.
In automatic motion calibrate the
speed
In Manual motion, move smoothly.
In automatic motion calibrate the
speed
In Manual motion, move smoothly.
In automatic motion calibrate the
speed
In Manual motion, move smoothly.
In automatic motion calibrate the
speed
In Manual motion, move smoothly.
In automatic motion calibrate the
speed
In Manual motion, move smoothly.
In automatic motion calibrate the
speed
In Manual motion, move smoothly.
In automatic motion calibrate the
speed
153