CEILING SUSPENSION
PROBLEM
CEILING SUSPENSION CAN NOT BE
SWITCHED ON
CEILING SUSPENSION ON,
CONTROL CONSOLE OFF
WRONG DISPLAY MEASURES
8.4
SOFTWARE ERROR MESSAGES
8.4.1
ERROR DISPLAY
Illustration 8-1
Error Display
1
2
4
First configuration
142
CHECK IF
Emergency Stop Switch is activated.
There is not enough power.
Check
Control
Console
connections.
Wrong calibration.
The ERROR MESSAGE is displayed in the Control Console.
1.
Error Message.
2.
Cause of the Error.
3.
Action to be completed to avoid the Error. This last Level is not displayed
with all errors.
4.
Press the Error Button to close the Error Message Display.
There are two different display configurations.
Deactivate Emergency Stop Switch.
Check that the Line Power is provided to the
Ceiling Suspension from the Room Electrical
Cabinet. If it is correct but it can not be turned ON,
contact Service.
Cables
Contact Service.
Contact Service.
1
2
3
4
Second configuration
ACTION
0242F EN 20181218