AGFA DX-D 600 User Manual page 160

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

DX-D 600
User Manual
Table 8-5 (Cont.)
Positioning
CODE
USER ERROR MESSAGE
Operative rejected: System
11521
synchronizing, please wait and try again
Operative rejected: System
11522
synchronizing, please wait and try again
Operative rejected: Please, release
11523
the MANUAL activators to proceed
Auto-center SKIPPED: Positioner is
11524
already centered
Operative rejected: Tracker not
allowed for DIRECT workstation:
11525
Please, check WORKSTATION
selection
Auto-tracking forbidden: System
11526
synchronizing, please wait and try
Auto-tracking forbidden: System in
11527
motion, please wait and try again
Operative rejected: Please, release
11528
the MANUAL activators to proceed
Operative rejected: The Detector and
11529
the Tube-crane are misaligned:
Please, get/set a valid SID
Auto-tracking skipped: Function
11530
already activated
Operative rejected: System
11531
synchronizing, please wait and try again
Operative rejected: System
11532
synchronizing, please wait and try again
Operative rejected: Please, release
11533
the MANUAL activators to proceed
Stitching PileUp: Positioner
11534
synchronizing, please wait and try
Auto-movement Failure: Positioner out
11535
of the auto-centering zone: Please, try
the operative again
Auto-movement aborted: Timeout to
11539
position the Detector: Please Try
Operative rejected: The
11540
auto-movement interlock is activated:
Please, check over-speed cause
152
Auto-tracking function is not available
again
The Bucky/Detector and the Tube are
again
again
ERROR DESCRIPTION
System synchronizing
System synchronizing
All brakes activated
Positioner is already centered
when DIRECT Workstation is
configured
System synchronizing
System in motion
All brakes activated
misaligned
Auto-tracking function already
activated
System synchronizing
System synchronizing
All brakes activated
ACTION
Wait until synchronizing procedure
finishes
Wait for system stop
Press any Control Console brake
button
Try again
Change WORKSTATION
Wait until synchronizing procedure
finishes
Wait for system stop
Press any Control Console brake
button
Align X-ray Tube and Bucky/Detector
Try again
Wait until synchronizing procedure
finishes
Wait for system stop
Press any Control Console brake
button
Try again
Try again
Try again
Check the over-speed cause
0242F EN 20181218

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents