Table 8-3 (Cont.)
Wall Stand Motor Controller Boards
CODE
USER ERROR MESSAGE
End-switch limit:
11212
Wall Stand-Vertical
Position out of range:
11213
Wall Stand-Tilting
11214
Control Board Error: Wall Stand-Tilting
11215
End-switch limit: Wall Stand-Tilting
0242F EN 20181218
ERROR DESCRIPTION
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
DX-D 600
User Manual
ACTION
Check End Switch Limit position,
activation, operation.
Check End Switch Limit range.
Check Axis calibration
Replace Control Driver Board
Check End Switch Limit position,
activation, operation.
Check End Switch Limit range
149