4.2.3.6
AUTO-CENTER
.
Note
.
Note
.
Note
.
Note
0242F EN 20181218
The Ceiling Suspension gets aligned the X-ray Tube with the center of the
DR Detectors of the RAD Table or RAD Wall Stand, it is not active with the
Direct Workstation.
DUE TO SAFETY REASONS, THE MINIMUM SID FOR
AUTO-CENTER FUNCTION IS HELD AT 950 mm (37.4") FROM
THE DR DETECTORS.
Automatic Collimators should change to Manual mode if the
Ceiling Suspension and the DR Detector are not at 2
orthogonality.
Auto-center movement is initiated by the Automatic Positioning
Control Box. For safety reasons, press and hold the Auto-center
button to maintain the motion, once it is released the motion stops.
Auto-center is paused as soon as the brake of the selected
DR Detector is pressed. Once the DR Detector brake button is
released, the auto-centering movement continues.
Auto-center is aborted as soon as the brake of the non selected
DR Detector is pressed. In this case it is necessary to restart
auto-center.
If the selected DR Detector position is changed during the
auto-center movement, this will be recalculated after the DR
Detector brake is released. Then, the movement is restarted
automatically to its new final centering stop.
To complete the Auto-center function:
1.
Select the Workstation.
2.
Configure in POSITION the desired SID. Refer to Section 4.2.4 for
Automatic System or Section 4.2.5 for Semi-automatic System.
3.
Press on the Touchscreen Console the AUTO-CENTER Control to
activate the function.
DX-D 600
User Manual
o
of the
81